Home / Scientific activities / Researches / A New Design Micro Total Analysis System for Determination Hydrazine in Pharmaceutical Samples

A New Design Micro Total Analysis System for Determination Hydrazine in Pharmaceutical Samples

Muthana Saleh Mashkoor*, Ahmed Adnan Al-Khafagi* , Dakhil Nassir Taha**
* Chemistry department-Faculty of Science /Kufa University, Iraq
** Chemistry department-Faculty of Science / Babylon University, Iraq.
Abstract
The new microfluidic analysis system use to determine the hydrazine based on the reaction between Spectrophotometric reagent vanillin with sulfuric acid and hydrazine
sulfate to form the hydrazone complex . Wavelength 410 nm , and give a linear range (7.685 x 10
-6 – 5.379 x 10-4) molar .The detection limit is 7.685 x 10-6 molar and the
regression coefficient R
2 value is 0.9979. Reproducibility, SD, RSD, dead volume and interferences have been calculated. The reaction time, concentration of reagent and
volume of sample has been optimized. The method has been successfully applied for the determination of hydrazine in water and pharmaceutical samples.
Key Words : Microfluidic analysis, Chip , Hydrazine , Vanillin , pharmaceutical samples.

Check Also

Serum Levels of lead, cadmium and silver in Patients with Breast Cancer compared with healthy females in Iraq

Shaymaa Awad Kadhim Alshebly1, a, Hayder H. Hussain2, b, Saleem Hamza Trier3, c Baydaa Awad …